Disintegration test
for sugar-coated tablets
Procedure:
1.
The
apparatus was set up for the disintegration test according to it’s operation
manual.
2.
The
temperature of the disintegration medium (water) was ensured at 37±2°C.
3.
The
time was set to 60 minutes. One Antibiotic tablet was introduced into each
tube, the disk was added into each tube and the operation was started.
4.
The
Antibiotic tablet was checked in each tube at the end of the operation.
5.
The
Antibiotic tablet was complied with the test if it is disintegrate in 60
minutes. If there is any Antibiotic tablet that does not disintegrate, the test
was repeated using the same tablet but the disintegration medium (water) was
replaced with 0.1M hydrochloric acid. The Antibiotic tablet was complied with
the test if it is disintegrate in the acidic medium.
Results:
Tablets
|
Disintegration test
with water
|
Time needed to fully
disintegrate
|
1
|
Pass
|
2.23 minutes
|
Discussion:
Disintegration
test is only applicable for uncoated tablets, sugar coated tablets ,
enteric-coated tablets, soluble tablets, dispersible tablets, hard capsules,
soft capsules and enteric capsules. Antibiotic tablet, which was used in
this experiment, is a sugar coated tablets.
From this
test, in average, an Antibiotic tablet would need 2.23 minutes to be fully
disintegrate using Apparatus A setup. The standard set for this experiment was
to have the tablet disintegrate not more than one hour in water medium. So,
Antibiotic tablet have succeeded this test. But this results did not really
imitate how the preparation would disintegrate in human body. Multiple
parameters to really imitate our body system upon drug intake were not
provided. Some of these parameters include the different pH and motility in
different parts of gastrointestinal tract. This result is not guarantee to be
the same as in an actual person. It might be lot more faster since the motility
of human GI tract were quite vigorous and the pH were lower. There
might be preparation that failed this test but actually would be able to be
fully disintegrate in an actual human. Still, this result is still come in
handy in preformulation test so to reduce drug candidates thus to save time and
money.
Some
precaution need to be taken during this experiment. The coating of a
tablet did not subjected to be disintegrate within the time limit. We must pay
close attention on identifying that the content of the tablet comply with the
standard stated. When a formulation have reached the time limit but still have
bit of it that did not fully disintegrate, check by squishing it to see whether
the fragments still have part that did not come in contact with the water
(still in it dry hard tablet or dry powdered tablet form). If the first batch
did not meet the requirement, the test need to be repeated using same number of
new tablet but the medium now is 0.1M hydrochloric acid. Usually, preparation
that must succeed dissolution test did not have to do this test.
Conclusion:
All
Antibiotic tablets succeeded this test with the first tablet disintegrate in
2.23 minutes. This disintegration duration might not be the same as in real
human body. Most likely, it would need shorter time due to more harsh condition
(compared to subjected condition in the experiment).
Dissolution test for
tablets
Procedure:
1.
The
dissolution vessel was filled up with the buffer solution to 900 ml mark. The
temperature was set to 37°C.
2.
The
temperature of the dissolution medium was checked. It was ensured at 37±0.5°C.
3.
One
Ibuprofen Tablet of 200 mg was placed into each dry basket assembly.
4.
The
stirring speed was set to 150 rpm. The basket was lowered assembly into
position in the vessel and the operation was started.
5.
After
30 minutes, 10 ml samples of the dissolution medium were withdrawn from each
vessel for analysis and the solution was filtered using suitable filter. The
sampling was taken out from a point half-way between the surface of the
dissolution medium and the top of the rotating basket, and not less than 10 mm
from the wall of the vessel. The volume of aliquot withdrawn was replaced for
analysis with an equal volume of same dissolution medium.
6.
A
standard solution of ibuprofen was prepared by diluting 10.0 mg of ibuprofen reference
standard to 50 ml with dissolution medium.
7.
2.0
ml of sample solution and 2.0 ml of standard solution were diluted to 25 ml
with dissolution medium in separate volumetric flasks.
8. The absorption of both
solutions was measured in a 1 cm cell at a wavelength of 221 nm.
Results:
Type of solution
|
Spectrophotometer
absorbance reading
|
Standard
|
3.1351
|
Sample
|
0.7552
|
% amount of ibuprofen
dissolved
= 0.7552/3.1351 x
10mg/50 x 2/25 x 0.98 x 900 x 25/2 x 100/200
=21.25%
·
The
ibuprofen tablet is not complied with
the USP requirements
Discussion:
The
reading of spectrophotometer for standard solution and sample solution are
3.1351 and 0.7752 respectively. Based on the result, the percentage amount of
ibuprofen dissolved is 21.25%.
Thus, we
can conclude that this tablet is not complied with the requirements United
States Pharcomacopoiea as the percentage must be more than 75% to be approved
as the complied one.
There may
be several errors those have been made during the experiment is being
conducted. The dissolution process of ibuprofen tablet could be incomplete
thus, the spectrophotometer reading may be altered as spectrophotometer.
Besides, the sample solution is not filtered by using the filter when the
solution is being transferred into the volumetric flask. Thus, the small,
undissolved ibuprofen tablets are also present in the volumetric flask. This
will result in wrong reading of absorbance by the spectrophotometer due to the
wrong concentration of the solution. This is because the amount of light
absorbed by the solution is proportional to the concentration of the solute
(ibuprofen) present.
Furthermore,
the result obtained are not complied with the USP requirements maybe due to the
expired tablet. An expired tablet will have a change in physicochemical
properties including in the dissolution aspect thus resulting in incorrect
result. The tablet must be rechecked and new batch should be produced to be
tested again.
The
parallax error also may occur during the reading of solution in the measuring
cylinder and volumetric flask. The incorrect volume of both solution will alter
the concentration of the solution, thus result in wrong reading of absorption
in spectrophothometer. In order to overcome this problem, the experimenter must
make sure that his eyes are perpendicular to the plane of measuring cylinder
and the calibration mark of volumetric flask.
Conclusion:
From the experiment, the tested ibuprofen tablet is not
complied with the USP requirements as only 21.25% of the tablet dissolved
during the experiment.
No comments:
Post a Comment